Abstract
Purpose:
Many sophisticated and expensive trainers have been developed to assist surgeons in learning basic laparoscopic skills. We developed an inexpensive trainer and evaluated its effectiveness.
Materials and Methods:
The webcam laparoscopic training device is composed of a webcam, cardboard box, desk lamp and home computer. This homemade trainer was evaluated against 2 commercially available systems, namely the video Pelvitrainer (Karl Storz Endoscopy, Culver City, California) and the dual mirror Simuview (Simulab Corp., Seattle, Washington). The Pelvitrainer consists of a fiberglass box, single lens optic laparoscope, fiberoptic light source, endoscopic camera and video monitor, while the Simuview trainer uses 2 offset, facing mirrors and an uncovered plastic box. A total of 42 participants without prior laparoscopic training were enrolled in the study and asked to execute 2 tasks, that is peg transfer and pattern cutting. Participants were randomly assigned to 6 groups with each group representing a different permutation of trainers to be used. The time required for participants to complete each task was recorded and differences in performance were calculated. Paired t tests, the Wilcoxon signed rank test and ANOVA were performed to analyze the statistical difference in performance times for all conditions.
Results:
Statistical analyses of the 2 tasks showed no significant difference for the video and webcam trainers. However, the mirror trainer gave significantly higher outcome values for tasks 1 and 2 compared to the video (p = 0.01 and <0.01) and webcam (p = 0.04 and <0.01, respectively) methods. ANOVA indicated no overall difference for tasks 1 and 2 across the orderings (p = 0.36 and 0.99, respectively). However, by attempt 3 the time required to complete the skill tests decreased significantly for all 3 trainers (each p <0.01).
Conclusions:
Our homemade webcam system is comparable in function to the more elaborate video trainer but superior to the dual mirror trainer. For novice laparoscopists we believe that the webcam system is an inexpensive and effective laparoscopic training device. Furthermore, the webcam system also allows instant recording and review of techniques.
References
- 1 : Training postgraduate urologists in laparoscopic surgery: the current challenge. J Urol2002; 167: 2135. Link, Google Scholar
- 2 : Improving operative performance using a laparoscopic hernia simulator. Am J Surg2001; 182: 725. Google Scholar
- 3 : Housestaff performance is improved by participation in a laparoscopic skills curriculum. Am J Surg2002; 184: 626. Google Scholar
- 4 : Using simulators to assess laparoscopic competence: ready for widespread use?. Surgery2004; 135: 28. Google Scholar
- 5 : Training in laparoscopic suturing skills using a new computer-based virtual reality simulator (MIST-VR) provides results comparable to those with an established pelvic trainer system. J Laparoendosc Adv Surg Tech A2002; 12: 167. Google Scholar
- 6 : A cheap laparoscopic surgery trainer. Ann R Coll Surg Engl1992; 74: 256. Google Scholar
- 7 : Home-made videoscopic trainer for operative laparoscopic surgery. Aust N Z J Surg1995; 65: 820. Google Scholar
- 8 : A computer-based laparoscopic skills assessment device differentiates experienced from novice laparoscopic surgeons. Surg Endosc2001; 15: 1085. Google Scholar
- 9 : Acquiring surgical skills. Br J Surg2000; 87: 28. Google Scholar
- 10 : Assessment of individual hand performance in box trainers compared to virtual reality trainers. Am Surg2003; 69: 1112. Google Scholar
- 11 : The transfer of basic skills learning in a laparoscopic simulator to the operating room. Surg Endosc2002; 16: 1324. Google Scholar
- 12 : Evaluation of endoscopic and laparoscopic training practices in surgical residency programs. Surg Endosc2001; 15: 1011. Google Scholar
- 13 : Objective structured assessment of technical skill (OSATS) for surgical residents. Br J Surg1997; 84: 273. Google Scholar
- 14 : Evaluating minimally invasive surgery training using low-cost mechanical simulations. Surg Endosc2003; 17: 580. Google Scholar
From the Endourology Institute, Cedars-Sinai Medical Center (SYC, CHC, CSN, GJF), Los Angeles, California, and Biostatistics, University of Pittsburgh (DL), Pittsburgh, Pennsylvania